This is clearly stated in section 3.1.3.3 for use of expendable TC. My question is what's about use of nonexpendable base metal TCs for TUS?
In section 3.1.5.3, it was mentioned for the life of nonexpandable base metal load TCs. But nowhere mentioning the test TCs.
It's quite confusing.
EdStainess,
Thanks again for your reply.
Yes, it says type K is not allowed to be re-calibrated if used above 500F, but that's for re-calibrate rather than re-use. I am checking about the re-use of K type as test sensors. It should be different from re-calibration (As you can see, for using of...
EdStainless,
Thanks for your reply.
Non-expendable type K is not premitted for TUS testing at all? Is there any clear statement on this?
I am not quite understanding the idea behind this. As you can see, for recalibration, the expendable base metal sensor is not permitted while it is...
Any expert in familar with AMS 2750E could help to answer the specific question for me?
That is for the usage of non-expenable K type test sensors for TUS sensors.
In Figure 1, it is mentioned that non-expandable E&K type sensors are not permitted for re-calibration if used above 500F.
In...