Paolo55 is correct, VLE package detects very small amounts of heavy components as C9+, we split C6+ into C7 and C8 fractions to get results comparable with those of our laboratory instrument, results are reasonably accurate, a laboratory chilled mirror is required to verify regularly calculated...
we have two EOS based on GC composition systems and they match (1 C) our laboratory mirrors, in-line chilled mirrors have been found quite critical so we decided to calculate dew points in DCS, our GCs go up to C6 and we characterize C7-C8 with a simple splitting, results match those of our...